Chapter 8 Operating Instructions
CLV490 Bar Code Scanner
8-14 © SICK AG · Division Auto Ident · Germany · All rights reserved 8009993/RA61/2007-07-20
Troubleshooting
8.5.3 Malfunctions in Reading mode: result output errors
¾ Check whether the "Device Ready" LED lights up.
If the LED does not light, see Table 8-5, Page 8-9.
CLV490 in stand-alone configuration
Here: number of bar codes to be read = 1
Malfunction Possible cause Remedy
1. The "Read Result" LED (default setting:
Good Read) does not light up at the end
of the reading pulse.
The "Result 2" switching output
(default setting: Good Read) is not
outputting a pulse
The CLV490 outputs status ST=2 in
the reading result on the host inter-
face
(outputting of ST disabled by default)
The CLV490 outputs the message
"no code" in the reading result of the
auxiliary interface
Read not successful, since there was no
bar code in the reading field during the
reading pulse.
Scan line positioned incorrectly.
Bar code presented at incorrect reading
distance.
The presented bar code’s module width
cannot be read by the CLV490 type be-
ing used.
Reading range (bar code evaluation range):
Evaluation range of the scan line is
limited unfavourable (reading angle valu-
es).
or
Parametrized rectangular reading area
does not match the conditions in situ.
Position of the CLV490 referred to the
reading space not or incorrectly
parameterized.
Focus control:
Focus control not activated.
Synchronize read with presence of an
object in reading field. See Chapter 8.5.2
Malfunctions in Reading mode: reading
trigger errors, Page 8-10.
Align CLV490. Is bar code at center of
scan line? Check the reading quality with
the "CLV-Setup" program: choose P
ER-
CEN-
TAGE EVALUATION. See Chapter 6.5.2 Per-
centage evaluation, Page 6-41.
Check whether reading distance of bar
code is in the specified reading range?
Present a bar code with an acceptable
module width. See Chapter 9 Technical
data, Page 9-1.
In the "CLV-Setup" program: choose the
READING CONFIGURATION tab, click the AU-
TOFOCUS PARAMETERS respectively
DISTANCE CONFIGURATION button.
Are the min. and max. reading angle
values correct? Check with S
HOW RA-
LIMITS. See Chapter 6.5.5 Show RA-limits,
Page 6-46.
In the "CLV-Setup" program: choose the
READING CONFIGURATION tab, READING AREA
section, READING AREA option, click the
R
EADING SPACE PARAMETERS button: are
the values for reading space correct?
In the "CLV-Setup" program: choose the
D
EVICE CONFIGURATION tab, click the SCAN-
NER POSITION PARAMETERS button: are the
values correct?
In the "CLV-Setup" program:
choose the READING CONFIGURATION tab,
F
OCUS CONTROL section: "Autofocus" or
"Inputs/serial" for event-controlled focus
position switchover selected? (for oscil-
lating mirror also "Oscillating Mirror Ex-
trema")
Table 8-8: Troubleshooting: result output errors in Reading mode (CLV490 in stand-alone configuration)